Recurrent Inappropriate ICD Shocks Following Lead Fracture Due to Blunt Chest Trauma: A Case Report
Anshu Man Singh1, Shubham Srivastava1, Roopali Khanna1
1Sanjay Gandhi Postgraduate Institute of Medical Sciences.
Abstract:
Implantable cardioverter-defibrillators (ICDs) are effective in preventing sudden cardiac death; however, lead-related complications remain an important cause of device malfunction. Lead fracture may result in oversensing and inappropriate shocks, significantly affecting patient morbidity. We report a 48-year-old man with ischemic cardiomyopathy and a single-chamber ICD who presented with multiple inappropriate shocks following a motor vehicle accident. Device interrogation demonstrated high-frequency non-physiological noise, abrupt reduction in lead impedance, and premature battery depletion. Fluoroscopy confirmed ventricular lead fracture. The fractured lead was abandoned, and a new ventricular lead was successfully implanted. Inappropriate shocks resolved completely following lead revision. Trauma-related lead fracture should be considered in patients presenting with inappropriate ICD therapies after blunt chest trauma. Prompt device interrogation and lead revision are essential for appropriate management.
Related Concept Videos
Flail Chest-II
Assessment:
1. Clinical Evaluation:
History:
Flail Chest-I
Flail chest is a severe and potentially life-threatening condition characterized by the fracture of three or more adjacent ribs in multiple places. It is most commonly caused by direct impacts and trauma, such as motor vehicle accidents or injuries from a steering wheel impact. It can also occur due to falls in elderly individuals with osteoporosis, or assaults involving sharp objects.
Pathophysiology
The pathophysiology of flail chest is complex, involving fractures of...
Cardiopulmonary Resuscitation III: AED Use
Cardiopulmonary Resuscitation IV: Pharmacological Management
Pneumothorax-I
Pneumothorax can be even further classified as spontaneous, traumatic, and tension pneumothorax.