Rui Gu1, Jiayu Zhang1, Chungen Qian2
1Department of Pediatrics, and Department of Nephrology, Children's Hospital, Zhejiang University School of Medicine, National Clinical Research Center for Children and Adolescents' Health and Diseases, Hangzhou, 310052, China; Zhejiang University School of Medicine, Hangzhou 310058, China.
Introduction:
Circulating anti-nephrin autoantibodies are being increasingly recognized in autoimmune podocytopathies, particularly minimal change disease (MCD) and focal segmental glomerulosclerosis (FSGS). However, circulating anti-nephrin reactivity is often low, and currently available assays differ substantially in design, antigen production, and analytical performance. This methodological heterogeneity highlights the need for standardized and quantitative assays for clinical application.
Methods:
Recombinant human nephrin ectodomain-coated magnetic microparticles were used to establish an automated chemiluminescence immunoassay (CLIA). Serum anti-nephrin IgG was measured in 121 children with biopsy-confirmed MCD (n=103) or primary FSGS (n=18), together with disease controls (genetically confirmed MCD/FSGS, IgA nephropathy, IgA vasculitis with nephritis, lupus nephritis, and ANCA-associated glomerulonephritis) and healthy individuals. Samples were tested in parallel by immunoprecipitation-Western blot (IP-WB). Antigen inhibition assays and cohort-based clinical analyses were performed.
Results:
The CLIA met predefined analytical performance criteria and showed substantial agreement with IP-WB, with an overall concordance of 86.8% and Cohen's kappa coefficient of 0.714. CLIA detected anti-nephrin IgG more frequently than IP-WB in patients with MCD or FSGS (41.3% vs. 28.1%), including additional low-level positive samples. Antigen inhibition assays showed dose-dependent inhibition in CLIA-positive samples. Among patients with baseline nephrotic-range proteinuria, antibody positivity was associated with greater proteinuria, lower serum albumin, and steroid-sensitive nephrotic syndrome. Patients with elevated anti-nephrin IgG levels had significantly shorter relapse-free survival. During complete remission, anti-nephrin IgG positivity was associated with relapse within three months, with relapse risk increasing according to antibody level.
Conclusions:
Together, our automated CLIA may help move anti-nephrin IgG testing from specialized research workflows toward standardized clinical application, supporting future use in disease stratification and longitudinal monitoring.
More Related Videos
07:20A Simple Cell-based Immunofluorescence Assay to Detect Autoantibody Against the N-Methyl-D-Aspartate (NMDA) Receptor in Blood
Published on: January 9, 2018
06:50A High-Throughput Electrochemiluminescence 7-Plex Assay Simultaneously Screening for Type 1 Diabetes and Multiple Autoimmune Diseases
Published on: May 29, 2020
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Nephrotic Syndrome II : Assessment and Medical Management