Evaluation of an Upper Arm Capillary Blood Collection Device Compared with Venous and Fingerstick Sampling
Stacy G Beal1,2, Anissa Naranjo1, Carlos Torres1
1LetsGetChecked Laboratories, Monrovia, CA, United States.
Background:
Upper arm capillary blood collection devices may expand access to laboratory testing outside traditional phlebotomy settings. We evaluated the analytical performance and patient preference of the ImPress, an investigational-use upper arm capillary blood collection device compared with venous and fingerstick sampling.
Methods:
Matched samples were collected from 43 healthy volunteers. Sixteen routine chemistry and lipid analytes were measured on standard laboratory platforms. Agreement was assessed using Deming regression and bias at medical decision points (MDPs), with acceptability defined by total allowable error (TEa). Hemolysis indices were compared across collection methods. Participants completed a survey.
Results:
A total of 315 specimens were analyzed. Thirteen of 16 analytes met predefined acceptance criteria for upper arm capillary compared with venous sampling. Mean bias at MDP ranged from -3.1% to 10% for venous vs upper arm capillary samples. Albumin, calcium, and aspartate aminotransferase (AST) did not meet acceptance criteria, primarily because all samples were within a tight normal range. All correlation coefficients (R, Deming regression) were ≥0.90, indicating linear relationships between collection modalities. Hemolysis indices for samples collected by the upper arm capillary device were well below assay interference thresholds. Mean pain scores (0 to 10 scale) were lowest for upper arm capillary (0.55) compared with venipuncture (2.61) and fingerstick (2.87). Most participants (74%) reported a preference for an upper arm capillary device.
Conclusions:
Upper arm capillary blood collection demonstrated strong analytical agreement with venous and fingerstick sampling for most routine analytes, low clinically significant hemolysis, and high patient acceptability. These findings support its potential use in decentralized and remote laboratory testing.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...