Neonatal automated oxygen titration algorithms: response characteristics and performance
Hylke H Salverda1,2, Christoph E Schwarz3,4, Andrew P Marshall1,5
1Menzies Institute for Medical Research, University of Tasmania, Hobart, Tasmania, Australia.
Objective:
Neonatal respiratory support devices now have algorithms for automated adjustment of oxygen concentration (fraction of inspired oxygen (FiO2)) in response to oxygen saturation (SpO2) disturbances, but their performance has not been systematically examined. We investigated the response of six available devices and algorithms to standardised and simulated SpO2 disturbances.
Design:
Bench-top study in neonatal centres.
Methods:
Devices under study received SpO2 input from a vital signs simulator at 1 Hz; device response was recorded as measured FiO2 in the respiratory circuit. Response to protracted hypoxaemia (SpO2=80%) and hyperoxaemia (SpO2=98%) was examined in several test conditions. In closed-loop testing, measured FiO2 was combined with abstracted patient data in an oxygenation simulator, generating novel SpO2 values input to the device. Device capability in keeping simulated SpO2 in target range (91-95%) was examined, as was FiO2 adjustment frequency and propensity for FiO2 oscillation.
Results:
The FiO2 trajectories in response to SpO2 disturbances differed considerably between devices. Maximum rate of sustained FiO2 escalation (ΔFiO2/min) during hypoxaemia was Avea: 0.30; Fabian: 0.20; Leoni: 0.20; SLE: 0.10; Sophie: 0.062; Vapotherm: 0.30. Corresponding values for FiO2 reduction during hyperoxaemia were 1.20, 0.20, 0.20, 0.13, 0.53 and 1.32. In closed-loop testing, average SpO2 target range time proportion was 51%, 52%, 55%, 77%, 64% and 53% for the six devices, respectively, with 1.9-5.1 automated FiO2 adjustments/min, in some cases with a prominent oscillatory pattern.
Conclusions:
Automated oxygen control devices exhibit considerable variation in their FiO2 response to SpO2 disturbances, with important consequences for SpO2 targeting potential and the pattern of FiO2 adjustment.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
05:45Delivery of In Vivo Acute Intermittent Hypoxia in Neonatal Rodents to Prime Subventricular Zone-derived Neural Progenitor Cell Cultures
Published on: November 2, 2015
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Requirements and Growth Patterns
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.