Atrioventricular block and optimal synchronous pacing on ECG in a single lead VDD pacing population
Giuseppe Mascia1, Luca Barca2, Andrea Minghini2
1Cardiovascular Disease Unit, IRCCS Ospedale Policlinico San Martino, Genoa, Italy; Department of Internal Medicine and Specialties (Di.M.I.), University of Genoa, Genoa, Italy.
Introduction:
In patients with atrioventricular (AV) block and preserved sinus node function on 12‑leads electrocardiogram (ECG) a single‑lead VDD pacing is today an established alternative to dual-chamber pacing, potentially offering reduced complications. Optimal pacing in AV block should maintain the AV synchrony on ECG in order to maximize cardiac output as well as stroke volume. Therefore in single‑lead VDD pacing population a proper planning of the AV interval is critical, as it allows the atrial contraction to complete before ventricular activation begins.
Methods And Results:
We enrolled 67 consecutive patients with complete AV block and preserved sinus node function, with a single‑lead VDD pacemaker indication. According to atrial dipole length patients were divided into a long-dipole group (>20 mm, n = 25) and a short-dipole group (≤20 mm, n = 26), while 16 patients were excluded (8 due to loss at follow-up, 8 for clinical non-eligibility considering atrial fibrillation and/or documented sinus node dysfunction). The mean atrial sensing amplitude was 0.60 ± 0.56 mV in the long-dipole group versus 1.16 ± 0.92 mV in the short-dipole group (p = 0.02) at 6-months follow-up. AV synchrony was significantly higher with long dipoles (95.4 ± 4.6%) compared with short dipoles (84.6 ± 25.5%; p = 0.04) at same follow-up.
Conclusions:
In a single‑lead VDD pacing population, a longer atrial dipole is associated with a superior AV synchrony on ECG despite lower atrial sensing, supporting the concept that detection stability and electro-geometric consistency may prevail over absolute signal amplitude in preserving AV synchrony.
More Related Videos
08:05Optimization of Transesophageal Atrial Pacing to Assess Atrial Fibrillation Susceptibility in Mice
Published on: June 29, 2022
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias VI: Management of Dysrhythmias
Electrophysiology of Normal Cardiac Rhythm
Dysrhythmias V: Evaluating Dysrhythmias