Related Experiment Video
Updated: Sep 29, 2026

Measurement of Spatial Stability in Precision Grip
Published on: June 4, 2020
Responsiveness of the K-Force Grip Dynamometer in patients with ICU-acquired weakness: a prospective longitudinal
Ingrid D Van Iperen1, Marleen Flim2, Peter E Spronk1
1Department of Intensive Care Medicine, Gelre Hospitals, Apeldoorn, The Netherlands; Expertise Center for Intensive Care Rehabilitation - ExpIRA, Apeldoorn, The Netherlands.
Objective:
Handgrip strength is commonly used to assess muscle strength in critically ill patients, including those with intensive care unit-acquired weakness. While the reliability and validity of the K-force grip dynamometer have been established, its responsiveness remains unknown. This study evaluated the responsiveness of handgrip strength measured with the K-force grip dynamometer in intensive care unit-acquired weakness patients.
Methods:
In this prospective longitudinal study, intensive care unit-acquired weakness patients were assessed at baseline and after 1 week. Responsiveness was examined using effect sizes and standardized response means. The Medical Research Council sum score was included as a clinical comparator of global measure strength. Changes over time were analysed using Wilcoxon signed-rank tests.
Results:
Thirty-six patients were included. Dominant handgrip strength increased from 2650 g to 3450 g, with an effect size of r = 0.86 and standardized response means of 0.77 (p < 0.001). Non-dominant handgrip strength increased from 2,700 g to 3,783 g, with an effect size of r = 0.90 and SRM of 0.70 (p < 0.001). Medical Research Council sum score increased from 26.5 to 30.0 (p < 0.001).
Conclusion:
The K-force grip dynamometer demonstrated strong responsiveness in patients with intensive care unit-acquired weakness, supporting its use for detecting short-term changes in handgrip strength in critically ill patients.