Continuous monitoring of endotracheal tube obstructions using naturally occurring pressure and flow oscillations
Ben Fabry1, Christian Kuster2, Roland C E Francis3
1Department of Physics, Friedrich-Alexander Universität Erlangen-Nürnberg, Erlangen, Germany. ben.fabry@fau.de.
Background:
The endotracheal tube resistance dominates the total airway resistance in most intubated patients. Mucus deposition and biofilm formation can rapidly increase tube resistance and thereby contribute to serious ventilatory impairments, including dynamic hyperinflation, intrinsic PEEP build-up, added work of breathing, and patient-ventilator asynchrony. During controlled mechanical ventilation, an increased tube resistance can be inferred from the difference between peak and plateau pressure, but this approach fails during pressure-supported spontaneous breathing.
Methods:
We present a method that estimates the linear and nonlinear components of tube resistance from naturally occurring airway pressure and flow fluctuations at the airway opening, without a tracheal pressure sensor and without imposed forced oscillations. This is achieved by solving the equation of motion using band-pass filtered airway pressure and flow signals. Band-pass filtering isolates the relevant resistive and inertive pressure losses across the tube by removing slow contributions from muscle pressure and lung elastance as well as high-frequency noise.
Results:
Across different ventilation modes, ventilator settings, tube diameters, filters, and breathing patterns, the method recovers both linear and nonlinear tube resistance parameters with < 10% error and < 2% bias. The method can also track progressive tube obstruction on a breath-by-breath basis.
Conclusions:
Continuous estimation of the endotracheal tube resistance from naturally occurring, ventilator-induced airway pressure and flow fluctuations enables breath-by-breath detection of tube obstruction and provides the parameters required for automatic tube resistance compensation during spontaneous breathing. This approach may improve patient safety, reduce patient-ventilator asynchrony, and facilitate weaning.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...