Multidimensional T 1 - T 2 Relaxation Imaging In vivo on a Portable 0.064 T MRI Scanner
Ella Wilczynski1, Kulam Najmudeen Magdoom1,2,3, Alexandru V Avram1
1Section on Quantitative Imaging and Tissue Sciences, Eunice Kennedy Shriver National Institute of Child Health and Human Development (NICHD), National Institutes of Health, Bethesda, MD, USA.
Purpose:
To demonstrate the feasibility of whole-brain multidimensional relaxation imaging on a portable 0.064 T MRI scanner.
Methods:
A two-dimensional inversion-recovery, fast spin-echo acquisition was used to jointly encode and relaxation. Joint relaxation distributions were reconstructed voxel-wise using marginal-distribution constrained optimization (MADCO). The approach was evaluated in a quantitative relaxation MRI phantom and subsequently applied in vivo in a healthy volunteer. Joint distributions, marginal distributions, distribution-derived measures, and statistical dependencies between the two relaxation dimensions were examined.
Results:
Phantom joint relaxation estimates showed good overall agreement with mono-exponential relaxation measurements and previously reported values at 0.064 T. The in vivo acquisition provided whole-brain coverage over a broad range of relaxation weightings and tissue contrasts. Reconstructed joint distributions showed spatially organized features across the relaxation space, including broad and overlapping relaxation components that were not fully represented by single-value relaxation maps or by either 1D marginal distribution alone. Selected regions of the joint relaxation space produced component maps with distinct spatial patterns. Statistical analysis further showed a dependence between the two relaxation dimensions, supporting the presence of added information in the joint distribution that is lost when treating and separately.
Conclusion:
Whole-brain multidimensional relaxation imaging is clinically feasible on a portable 0.064 T MRI scanner. Joint relaxation distributions provide information beyond single-value relaxation mapping and may support further developments of quantitative multidimensional imaging at ultra-low field.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic Stimulation (TMS).