Related Experiment Video
Updated: Oct 3, 2026

Applying X-ray Imaging Crystal Spectroscopy for Use as a High Temperature Plasma Diagnostic
Published on: August 25, 2016
Single-VIPA dual-mode absorption-emission spectroscopy for plasma diagnostics
Abstract:
Combined absorption-emission spectroscopy is demonstrated using a single VIPA spectrometer coupled to a supercontinuum source. The system captures a ~36 nm bandwidth (742-778 nm) at picometer resolution. In a low-pressure glow discharge, N2 emission calibration, O I 777 nm triplet absorption (2.2% consistency), and linewidth analysis distinguishing instrumental/Doppler broadening from Ar I self-absorption (broadening factors up to ~1.5) are performed. Joint retrieval of temperature (953 ± 49 K) and instrumental width, along with time-division absorption-emission measurements, is demonstrated. Long-term stability makes this approach suitable for quasi-steady plasmas.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Emission Spectroscopy: Lab
Atomic Emission Spectroscopy: Instrumentation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
UV–Vis Spectrometers
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview