Misconnection of Oxygen Tubing to the "Smart Cuff Manager" Causing Cuff Explosion: A Preventable Equipment Error
Alexander Tzabazis1, Christoph Niemeyer2, Markus Stuhr1
1From the Department of Anesthesia, Intensive Care, Rescue and Pain Medicine, BG Klinikum Hamburg, Hamburg, Germany.
Abstract:
We report a case of sudden airway bleeding and airway loss secondary to an accidental misconnection of oxygen tubing to a "Smart Cuff Manager" (TRACOE, TRACOE medical GmbH) instead of a "Humidotrach" (heat and moisture exchange (HME) filter; Humidotrach Pro Iso, Fahl Medizintechnik-Vertrieb GmbH).
More Related Videos
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Errors occurring during blood pressure monitoring
Several factors...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...