Monte Carlo investigation of anode-filter combinations on X-ray spectra in conventional radiology using PHITS
Alex Eclador Ngankem1,2, Joel Cebastien Guembou Shouop3, Francis Chamberlain Djoumessi Zamo4
1Pole de Recherche, d'Innovation et d'Entrepreneuriat (PRIE), Institut Universitaire de la Côte, P.O. Box: 3001, Douala, Cameroun. ngankem.eclador@myiuc.com.
Abstract:
This research project investigates conventional radiology protocols by using PHITS (Particle and Heavy Ion Transport code System) Monte Carlo simulations to evaluate anode-filter configurations at tube voltages up to 150 kV. The study's innovation lies in its quantitative comparison of simulated spectra with SpekCalc, showing a mean deviation of less than 5%. The results indicate that Tungsten (W) and Gold (Au) anodes (K-lines: 59-68 keV) are optimal for radiography, whereas Molybdenum (Mo) and Rhodium (Rh) anodes (K-lines: 17-23 keV) are superior for mammography. By quantifying the impact of heavy filtration on beam hardening, this work establishes a robust framework for improving image contrast and supporting dose optimisation through appropriate selection of tube parameters.
Related Concept Videos
X-ray Imaging
Atomic Emission Spectroscopy: Interference
Mass Analyzers: Common Types
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences