Effect of Personalized Approach to CT Scan Length on Radiation Dose in Parathyroid SPECT/CT
Anne H Thostrup1,2, Natalie A Bebbington3, Janne Frederiksen1
1Department of Nuclear Medicine, Clinical Cancer Research Center, Aalborg University Hospital, Aalborg, Denmark.
Abstract:
Parathyroid SPECT/CT is the standard examination for localization of hyperfunctioning parathyroid tissue before surgery for hyperparathyroidism. The effective dose (ED) delivered by CT comprises approximately one third of the radiation dose in this examination. Parathyroid SPECT/CT typically includes 1 CT covering the mediastinum because of the risk of ectopic adenomas. This study investigates whether the use of personalized CT (pCT) on the basis of SPECT images can reduce scan length and CT-derived ED, while maintaining the ability to detect ectopic adenomas, compared with full field-of-view (FOV) SPECT/CT. Methods: Patients with hyperparathyroidism who were referred for a clinical parathyroid SPECT/CT between May 2023 and June 2024 were prospectively included. Two observers (a technologist and junior physician) retrospectively determined pCT based on the non-attenuation-corrected SPECT images. The reduction in CT scan lengths compared with those of standard CT examinations was calculated. In a subgroup of patients, the reduction in ED was calculated. The detectability of ectopic adenomas was analyzed using imaging data from the full FOV SPECT/CT as the reference standard. Results: In total, 113 patients were included in the study. The standard examination found that 82 patients had at least 1 suspected lesion. Scan length was reduced by 60%-66% using pCT. No ectopic adenomas were missed on pCT. Subgroup analysis showed comparable reductions in CT scan length and CT-derived ED. Conclusion: Median CT scan length and median CT-derived ED was reduced by 60% when pCT was used for parathyroid SPECT/CT. pCT was implemented with minimal delay between SPECT and CT.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...