DEVELOPMENT AND PRELIMINARY BENCHTOP EVALUATION OF A NOVEL ATHERMAL SEPTOSTOMY CATHETER
Majid Jadidi1, Jason MacTaggart2, Bahman Kargarbahrkhazar1
1University of Nebraska Omaha, Omaha, NE.
Abstract:
Chronic Type B aortic dissection (cTBAD) often requires aortic septostomy to depressurize the false lumen and enable effective endovascular repair. However, current "off-label" techniques, such as mechanical "cheese-wire" or electrosurgical cutting, lack precision and carry significant risks of uncontrolled tearing and collateral thermal injury. To address this unmet need, we developed the Aquablade catheter, which utilizes a dual-guidewire system and a contained, athermal, high-velocity fluid jet to precisely cut the aortic septum longitudinally. This study reports the design and preliminary ex vivo benchtop evaluation of the device. Functional prototypes were tested on human aorta and standard vascular graft materials in a custom benchtop setup. The device successfully created clean, continuous longitudinal cuts through both biological tissue and synthetic graft material. Visual inspection confirmed that the device's deflector anvil protected off target areas from injury. These results validate the Aquablade's ability to perform precise, athermal septostomy without collateral damage. This technology offers a promising, safer alternative for managing complex chronic dissections, supporting further evaluation in preclinical models.
More Related Videos
06:48Rapid Formation and Testing of Self-expanding NiTi Frames with a Small Form Factor Suitable for Minimally Invasive Implants
Published on: March 7, 2025
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015