Related Experiment Video
Updated: Oct 10, 2026

Low-Dose Gamma Radiation Sterilization for Decellularized Tracheal Grafts
Published on: April 14, 2023
Is it safe to use surgical instruments with debris from a previous surgery after sterilization?
Daniyil Semeshchenko1,2, Alicia Farinati2,3, Iván Alfredo Huespe4
1'Sir John Charnley' Hip Surgery Unit, Institute of Orthopaedics 'Carlos E. Ottolenghi', Hospital Italiano de Buenos Aires, Buenos Aires, Argentina.
Aims:
The presence of biological debris on surgical instruments after reprocessing poses potential risks of infection in orthopaedic surgery. While autoclave sterilization is the standard method, its effectiveness in eliminating microbial contamination from residual bone, cement, and fatty tissue remains insufficiently studied. The primary objective was to study in vitro the sterility of retained cement, bone, and fat tissue debris on surgical instruments after autoclave sterilization. The secondary objective was to compare, based on colony-forming units (CFU), which material exhibits greater biofilm formation.
Methods:
An in vitro study was conducted to evaluate the sterility of retained biological debris on surgical instruments after autoclave sterilization. The autoclave sterilization procedure was performed at 1 atmosphere pressure (121°C) for 20 minutes. A total of 350 repetitions were performed, including five instruments used as sample specimens (drill bit, cannulated flexible reamer, femoral broach, cannulated screwdriver, and wire passer), with 70 repetitions per specimen. CFU/mL were counted and biofilm morphology was assessed through scanning electron microscopy (SEM).
Results:
After autoclave sterilization, no positive cultures were detected in any of the 350 repetitions (0%, 95% CI 0 to 5; n = 0), demonstrating complete absence of viable microorganisms. Due to the lack of microbial recovery, no statistical comparisons between instrument types were performed. We did not find any structural deformity in the mixed biofilms on bone, cement, and stainless steel as observed by SEM.
Conclusion:
Autoclave sterilization was effective in eliminating microorganisms from surgical instruments with retained biological debris. Although no apparent structural deformities of microorganisms were observed by SEM in the mixed biofilms, no bacterial growth was detected after sterilization, supporting the microbiological effectiveness of autoclave sterilization even in the presence of retained biological debris. These findings provide microbiological evidence supporting current sterilization practices, and highlight the need for further studies evaluating alternative sterilization methods such as ethylene oxide or plasma.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Hand hygiene
Hand washing...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...