Related Experiment Video
Updated: Oct 13, 2026

Tissue-simulating Phantoms for Assessing Potential Near-infrared Fluorescence Imaging Applications in Breast Cancer Surgery
Published on: September 19, 2014
Probe-Based Near-Infrared Autofluorescence for Parathyroid Identification During Thyroid Surgery: Clinical Outcomes,
Anthony Y Cheung1,2, Bo Wang1,3, Yoshiyuki Saito1,4
1Division of Thyroid and Parathyroid Endocrine Surgery, Department of Otolaryngology-Head and Neck Surgery, Massachusetts Eye and Ear, Harvard Medical School, Boston, Massachusetts, USA.
Background:
We evaluated experience with probe-based near-infrared autofluorescence (NIRAF) system (PTeye) in parathyroid gland (PTG) identification during total thyroidectomy.
Methods:
In this retrospective study, patients undergoing total thyroidectomy using PTeye (2020-2022; n = 84) were compared with a historical visual-identification cohort (2018-2019; n = 103).
Results:
PTG identification rates were comparable, although 42% of PTGs detected with PTeye were initially visualized with low or moderate confidence. Significantly fewer patients in the PTeye cohort had PTGs incidentally identified on histology (4.8% vs. 16.5%, p = 0.018), but there was no difference in the incidence of autotransplanted PTGs or total inadvertent parathyroidectomy. Postoperative calcium was higher with PTeye (median 9.2 vs. 9.0 mg/dL, p < 0.001). Postoperative PTH, transient, and permanent hypoparathyroidism did not differ. An autofluorescence ratio cutoff of 2.5 improved specificity compared with the manufacturer-recommended cutoff of 1.2.
Conclusions:
PTeye provides adjunctive confirmation of suspected parathyroid tissue, including glands initially identified with low or moderate visual confidence.