Establishing the reliability of non-contrast 4D flow in congenital heart disease at 1.5T
Related Experiment Video
Updated: Oct 15, 2026

Human Fetal Blood Flow Quantification with Magnetic Resonance Imaging and Motion Compensation
Published on: January 7, 2021
Nicholas K Brown1,2, Erin K Romberg3, Christopher Schmitt4
1Department of Pediatrics, Division of Cardiology, Seattle Children' Hospital, 4800 Sand Point Way NE, Seattle, WA, 98105, USA. nkbrown2@texaschildrens.org.
Background:
4D flow has been shown to correlate well with 2D phase-contrast (2D PC) flow measurements, although most of the current literature is based on post-contrast imaging.
Objective:
Given recent interest in non-contrast techniques, we hypothesized that non-contrast 4D flow (NC 4D Flow) would correlate well with 2D PC and ventricular stroke volumes.
Methods And Materials:
A prospective cohort of pediatric and young adult patients with suspected/confirmed congenital heart disease (CHD) underwent cardiac MRI at 1.5T with acquisition of NC 4D Flow and 2D PC. Pearson correlation, linear mixed-effect correlations, and Bland-Altman comparisons were performed to compare these methods with secondary analyses of internal consistency and image quality/noise.
Results:
In our cohort of 30 patients (age, 8-20 years), NC 4D Flow correlated strongly with 2D PC for net flow (r=0.76), peak velocity (r=0.9, P<0.001), and regurgitant fraction (r=0.75, P<0.001). By Bland-Altman comparison, NC 4D Flow was 10.1% lower for net flows and 18.9% higher for peak velocity compared to 2D PC. NC 4D Flow also slightly underestimated ventricular stroke volumes (18.1% lower for ascending aorta vs. left ventricular stroke volume and 13.9% lower for main pulmonary artery vs. right ventricular stroke volume). Both datasets were internally consistent; 2D PC had a higher relative signal intensity, but no difference in signal-to-noise or contrast-to-noise ratios (SNR/CNR).
Conclusion:
Non-contrast 4D flow correlates well with 2D PC in CHD with a slight underestimation in flows and overestimation in peak velocities, but good internal consistency and comparable SNR/CNR.
More Related Videos
11:16In vitro Assessment of Aortic Regurgitation Using Four-Dimensional Flow Magnetic Resonance Imaging
Published on: February 25, 2022
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Magnetic Resonance Imaging