Comparison of two respiratory oscillometers in patients with a broad range of lung function patterns
Murphy Binnie1, Joyce K Y Wu1,2, Chung-Wai Chow1,2,3
1Division of Respirology, Department of Medicine, Temerty Faculty of Medicine, University of Toronto, Toronto, ON, Canada.
Background:
Oscillometry provides a non-invasive assessment of respiratory system mechanics and has shown potential for early detection of lung disease. However, differences in device hardware and signal processing may limit comparisons between devices. The objective of this study was to evaluate agreement between two commercial forced oscillometry devices, Tremoflo C-100 (Thorasys Medical Systems, Canada) and Resmon Pro (Restech srl, Italy) in patients with a spectrum of abnormal lung physiology.
Methods:
49 patients with chronic lung disease and 27 healthy volunteers completed paired oscillometry using both devices in an alternating order. Measurements of resistance at 5 Hz (R5), resistance between 5 and 19 Hz (R5-19), reactance at 5 Hz (X5), the resonant frequency of the lungs (Fres), and area of reactance (AX) were compared using Wilcoxon signed-rank tests, Bland-Altman analysis, intraclass correlation coefficients (ICC), and effect size calculations.
Results:
Significant differences were found between devices for all impedance measurements (p ≤ 0.001). The Resmon Pro reported lower R5, R5-19, Fres, and AX, and less negative X5 than the Tremoflo C-100. ICC estimates ranged from 0.59 to 0.85, but Bland-Altman analysis demonstrated systematic bias and wide limits of agreement; for AX, the bias was -4.10 cmH₂O/L with limits of agreement from -14.99 to 6.80 cmH₂O/L.
Conclusion:
Systematic biases and wide limits of agreement indicate that the devices should not be used interchangeably for longitudinal patient monitoring. Standardization of oscillometry device signal generation and processing is necessary to enable accurate cross-device comparison in both clinical and research applications.
Related Concept Videos
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration can...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes include...
Respiratory Volumes and Capacities I