Validating a Wearable Bluetooth Oximeter for Infants in the Neonatal Intensive Care Unit
Amy Courtney1, Ju Lee Oei1,2,3
1Faculty of Medicine and Health, University of New South Wales, Sydney, New South Wales, Australia.
Introduction:
Oxygen is vital for life but the wrong amount of oxygen can lead to significant risk of morbidity and mortality. Preterm infants are at particular risk from both too little (hypoxia) and excess (hyperoxia) oxygen. In the neonatal intensive care unit (NICU), oxygen treatment is guided by oximetry derived from wired oximeters which are difficult to use outside of the hospital environment. We aimed to validate Bluetooth wireless oximeters in newborn infants to determine the accuracy of devices that would be more suitable for out of hospital use.
Methods:
Three 20 min sets of measurements including oxygen saturations (SpO2), heart rate (HR) and perfusion index (PI) from the post-ductal regions (left arm and lower limbs) were conducted in 21 spontaneously ventilating infants using the Masimo (1) SafetyNet and (2) Radical 7 oximeters. The wired Radical 7 oximeter was the reference. Data were downloaded onto CSV files and analysed for agreement.
Results:
Twenty-one (eight male) infants were recruited between July to September 2023. Median gestations and ages were 36 weeks and 3 days. Neither Bluetooth device (SafetyNet and wireless Radical 7) met International Organization for Standardization (ISO) standard (ARMSE ≤ 3%), Score 3.624 (wireless Rad-7) and 3.884 (SafetyNet). Both wearables consistently overestimated SpO2, with mean biases (wired-wearable) of -1.802 (Rad-7) and -1.803 (SafetyNet), that increased at lower SpO2 levels. Linear regression, ICC, Pearson's correlation analyses showed poor consistency.
Conclusion:
Wireless oximeters have poor correlation with wired oximeters especially at lower SpO2 levels but ease of use, especially in the home environment in less acutely ill patients, requires further evaluation.
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...