Related Experiment Video
Updated: Oct 19, 2026

Polarization-Sensitive Two-Photon Microscopy for a Label-Free Amyloid Structural Characterization
Published on: September 8, 2023
Correction method and tissue simulation approach for system polarization mode dispersion in polarization-sensitive
Tatiana Soldati1, Patrick J González1, Mathilde A E Binnekamp1
1LaserLaB, Department of Physics and Astronomy, Vrije Universiteit, Amsterdam, Netherlands.
Abstract:
Polarization mode dispersion (PMD) caused by single-mode fibers and circulators is a major source of noise in fiber-based PS-OCT systems. It introduces an unknown wavelength dependent change of the polarization state incident on the sample, which can lead to erroneous local optic axis and retardation measurements. In this paper, we present a correction method for PMD and other possible wavelength dependent system distortions, based on a Jones and Stokes eigenvector analysis. The method was validated in a new PS-OCT simulation framework of birefringent tissue and system PMD. It was then used to correct in vivo lung data acquired with a depth-multiplexed PS-OCT system used in combination with a distal scanning endoscope. The method accurately retrieves the system PMD and the correct local simulated tissue retardation and relative optic axis in simulations. In in vivo lung data, the correction method yielded strong improvements in the uniformity of the optic axis image and reconstruction of birefringent layers, such as the airway smooth muscle and the perichondrium.