The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve
Published on: September 7, 2022
Fluoroscopy Computer Navigation Versus Manual Fluoroscopic Acetabular Positioning in Direct Anterior Total Hip
Karlos E Zepeda1, Carmelo Burgio1, Shrey Vachhani1
1Adult Reconstruction and Joint Replacement Service, Department of Orthopaedic Surgery, Hospital for Special Surgery, New York, NY, USA.
Background:
Fluoroscopy is commonly leveraged in direct anterior approach total hip arthroplasty (DAA-THA). Computer-navigation systems provide real-time guidance aiming to enhance component positioning, but their learning curves and component accuracy remains incompletely characterized. This study evaluated the learning curve and safe zone compliance after implementation of computer navigation in DAA-THA.
Methods:
This retrospective cohort study analyzed 200 consecutive primary DAA-THA cases performed by a single high-volume surgeon: 100 manual and 100 computer-navigated cases using the Velys Hip Navigation system. Acetabular component inclination and anteversion were measured on postoperative standing radiographs. Safe zone compliance was assessed using surgeon-defined, Lewinnek, and Callanan criteria. Learning curve analysis utilized segmented linear regression and cumulative sum methods. Operative time, fluoroscopy exposure, and component safe zone compliance were compared between groups.
Results:
Operative time stabilized after 6 cases (P < .05), while significant breakpoint was identified for radiation exposure Computer navigated cases had shorter operative time (76.1 ± 16.4 vs 82.3 ± 17.0 minutes, P < .05) without increased fluoroscopy exposure (28.3 ± 6.6 vs 28.8 ± 9.8 seconds, P = .25). Safe zone compliance was significantly higher with navigation for surgeon-defined (98.9% vs 91.6%, P = .04), Lewinnek (86.3% vs 68.4%, P = .01), and Callanan criteria (68.4% vs 49.5%, P = .01).
Conclusions:
Fluoroscopic based computer navigation was associated with higher safe zone compliance in DAA-THA with a rapid 6-case learning curve and reduced operative time, without increasing radiation exposure.