Ten-year Performance of the Multi-Layer Insulated Defibrillation Lead in the Medicare Population
Robert D Schaller1, Haris Haqqani2, Ulrika Birgersdotter-Green3
1Section of Cardiac Electrophysiology, Cardiovascular Division, Department of Medicine, Hospital of the University of Pennsylvania, Philadelphia, PA.
Background:
Long-term surveillance of implantable cardioverter-defibrillator (ICD) leads is essential because clinically important mechanical failures may emerge years after implantation.
Objective:
To evaluate 10-year mechanical complication-free survival of the Optisure ICD lead (Abbott, Sylmar, CA) in a large real-world Medicare population.
Methods:
Manufacturer device-tracking records were linked to Medicare Fee-for-Service claims and device complaint data using the EP PASSION framework. Patients receiving an Optisure lead from February 2014 through December 2024 were included. Mechanical lead-related complications were defined by claims containing both a diagnosis of mechanical lead malfunction and a lead-related surgical intervention more than 30 days after implantation. Kaplan-Meier survival was estimated under a best-case scenario including confirmed Optisure-related events and a worst-case scenario also including potentially related events.
Results:
The cohort included 26,830 patients with 27,033 Optisure leads. Mean age was 73.8 ± 8.6 years, 28.4% were women, and 49.8% had cardiac resynchronization therapy ICDs. Overall, 379 complications were classified as confirmed lead-related and 606 as confirmed or potentially lead-related. At 10 years, complication-free lead survival was 97.2% (95% confidence interval [CI], 96.7%-97.7%) in the best-case scenario and 95.6% (95% CI, 94.9%-96.2%) in the worst-case scenario. Survival did not differ significantly by age, lead length, coil configuration, or device type.
Conclusion:
The Optisure ICD lead demonstrated favorable mechanical performance through 10 years in routine clinical practice. These findings support both the long-term durability of Optisure and the use of integrated real-world data for post-market device surveillance.
More Related Videos
10:05Testing the Efficacy of Pharmacological Agents in a Pericardial Target Delivery Model in the Swine
Published on: July 7, 2016
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Cardiomyopathy II: Dilated Cardiomyopathy