One-Sample Calibration Curve on GC-MS Using Silicon Isotopologue Channels from Silylation Derivatization: Validation
Haiguang Tan1, Lexuan Yang1, Yunzhe Li1
1College of Biotechnology, Tianjin University of Science and Technology ;Key Laboratory of Industrial Fermentation Microbiology, Ministry of Education; Tianjin Key Laboratory of Industrial Microbiology; State Key Laboratory of Food Nutrition and Safety, Tianjin300457, P. R. China.
Abstract:
Quantitative GC-MS analysis of nonvolatile metabolites conventionally requires multisample external calibration, imposing substantial preparative overhead in high-throughput settings. Here, the integration of silylation derivatization with an isotopologue-based one-sample calibration curve (OSCC) on GC-MS platforms is reported. Trimethylsilylation introduces silicon atoms with invariant natural isotope abundances (28Si:29Si:30Si = 92.23%:4.68%:3.09%), enabling multiple calibration anchors from a single standard injection. Two strategies were developed: OSCC with multiple isotopologue selected ion monitoring (OSCC-MISM) on GC-MS, serving as a compositional-verification precursor, and OSCC with multiple isotopologue reaction monitoring (OSCC-MIRM) on GC-MS/MS, recommended for real-sample quantification. Both were validated against multisample calibration curve (MSCC)-SIM and MSCC-MRM across 12 nonvolatile organic acids, achieving R2 > 0.99, RSDs below 16.1%, and recoveries of 80.5-111.2%. In 15 commercial Baijiu samples, Bland-Altman analysis of 180 paired measurements gave a pooled bias of +0.8%, with 95% limits of agreement within the prespecified ±15% margin, supporting equivalence of OSCC-MIRM with MSCC-MRM.
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...