Related Experiment Video
Updated: Oct 25, 2026

Simultaneous fMRI and Electrophysiology in the Rodent Brain
Published on: August 19, 2010
An MRI-compatible electrode implantation system for linking behavioral neurophysiology and stimulation-evoked fMRI in
Ryo Ito1, Zhiqing Wan2, Ken-Ichi Amemori1
1Institute for the Advanced Study of Human Biology (WPI-ASHBi), Kyoto University, Yoshida-Konoe-cho, Sakyo-ku, Kyoto 606-8501, Japan.
Abstract:
Local electrophysiology and microstimulation reveal site-specific neural activity and physiological effects but do not directly link these local effects to whole-brain responses from the same target site. We developed an MRI-compatible electrode implantation system for macaques that enables CT/MRI-based localization, postoperative depth adjustment, neural recording, microstimulation during behavioral testing, and subsequent electrical microstimulation combined with functional MRI (EM-fMRI) using the same implanted electrode. The system supported local field potential recording, a stimulation-induced reduction in pupil size, and distributed stimulation-evoked blood oxygenation level-dependent (BOLD) responses. This approach provides a practical framework for linking local neural activity and stimulation-induced physiological responses with whole-brain responses at the same target site.