The Quantification of Injectability by Mechanical Testing
Published on: May 13, 2020
Attributes Affecting Rigid Needle Shield Pull-Off Force and Cap Removal Force in Autoinjector Systems
Emran Lallow1, Zhaoxi Zheng2, Michasel Galluppi2
1Device Development and Technology, Merck & Co., Inc., Rahway, NJ 07065, United States emran.lallow@merck.com guangli.hu@merck.com.
Abstract:
Autoinjector cap removal force (CRF) must be controlled to support reliable cap removal and ease of use. To understand this force, syringe-level rigid needle shield (RNS) pull-off force (POF) must be considered together with the manufacturing and test attributes that can change the RNS interface before or during cap removal. This study evaluated how pull-off test speed, assembly-induced mechanical disturbance, applied axial preload, and elastomer siliconization affect POF and CRF. POF increased linearly with test speed at low-to-intermediate rates before reaching a plateau, showing that measured POF depends on the test rate. Autoinjector assembly reduced device-level CRF relative to syringe-level POF, while applied axial preload similarly reduced subsequent syringe-level POF, indicating that interface conditioning during assembly can alter the relationship between syringe-level POF and device-level CRF. Additional silicone oil modestly decreased POF, whereas one-month room-temperature aging after preload and 60°C/18h heat exposure after recapping each produced only a small additional increase under the evaluated conditions. Together, these results show that device-level CRF cannot be interpreted from syringe-level POF alone without accounting for test speed, assembly-induced interface conditioning, and material or aging effects.