Related Experiment Video
Updated: Oct 26, 2026

Phenol Red Thread-based Sampling Procedure for Untargeted Tear Fluid Lipidomics in Biomarker Discovery
Published on: December 12, 2025
Cascade recognition-enabled portable fluorescent sensor for specific detection of lysozyme in tear samples
Shanshan Wei1, Yuelin Yan1, Kaixin Hu2
1School of Chemistry and Materials Science, Weinan Normal University, The Western Part of the Middle Section of Chaoyang Street, Weinan, 714099, PR China.
Abstract:
Accurate detection of decreased tear lysozyme (LYZ) levels, a diagnostic biomarker for dry eye disease (DED), is critical for early intervention. Conventional LYZ detection requires bulky instruments; emerging aptamer/electrochemical methods are costly and complex. To address these challenges, we designed a portable fluorescent sensor using boron-doped red-emitting carbon dots (B-RCDs) via a three-step cascade recognition mechanism: electrostatic pre-enrichment, dynamic quenching, and boronate ester recognition. This cascade enables sensitive and selective detection of LYZ in human tear samples. Under optimal conditions, the sensor exhibits a linear response from 10 to 90 μg/mL with a LOD of 0.026 μg/mL, outperforming most reported CD-based probes. The sensor shows excellent anti-interference performance, with recoveries of 88.70%-106.30% (RSD < 4.0%) in real tear samples, consistent with standard turbidimetric results. Furthermore, a filter paper strip integrated with a smartphone-assisted color analysis module (Manhattan distance calibrated) achieves visual point-of-care LYZ detection within 5 min. This work provides a portable platform for point-of-care DED screening and a generalizable cascade recognition strategy for glycoprotein detection.