High-amplitude bubble CPAP versus a high-flow tracheal comparator in a rat model of acute lung injury
Chun-Shan Wu1, Yen-Kuang Lin2, Hsiu-Chu Chou3
1Department of Pediatrics, Taoyuan General Hospital, Ministry of Health and Welfare, Taoyuan, Taiwan, ROC.
Abstract:
Bubble continuous positive airway pressure (BCPAP) superimposes pressure oscillations on baseline distending pressure, but the incremental benefit of greater oscillation amplitude remains uncertain. We compared high-amplitude BCPAP with conventional BCPAP and a non-bubbling high-flow tracheal comparator in a rat model of acute lung injury induced by saline lavage followed by a 30-min ventilation-associated second mechanical insult. Rats were assigned to high-amplitude BCPAP (135° expiratory limb, n=8), conventional BCPAP (0°, planned n=4), or the high-flow tracheal comparator (n=8). Sequential waveform recordings in one additional rat showed oscillation amplitudes of 0.96 and 0.70cm H₂O and mean airway pressures of 5.35 and 5.08cm H₂O for 135° and 0°, respectively; these descriptive measurements were not subjected to inferential testing. Compared with the high-flow tracheal comparator, high-amplitude BCPAP was associated with higher PaO₂ and pH, lower PaCO₂, and lower lung wet-to-dry ratio, bronchoalveolar lavage fluid protein, TNF-α, and IL-6. Compared with conventional BCPAP, no sustained blood-gas advantage was detected; IL-6 was lower and respiratory rate was lower only at 140min. Histological injury scores did not differ among groups. These findings support a benefit of the high-amplitude BCPAP configuration over the high-flow tracheal comparator for selected outcomes, but do not establish a consistent incremental benefit over conventional BCPAP or an effect attributable specifically to oscillation amplitude.