Ultrafast Time-resolved Near-IR Stimulated Raman Measurements of Functional π-conjugate Systems
Published on: February 10, 2020
Baseline correction using DU-GridNet for Raman spectra
YoungJae Son1, Guangyong Shang2, Tiejun Chen2
1Department of Intelligent Electronics and Computer Engineering, Chonnam National University, Jeonnam-Gwangju 61186, South Korea. tozero@jnu.ac.kr.
Abstract:
Raman spectroscopy is a pivotal tool for molecular and material characterization; however, fluorescence backgrounds often severely distort measured signals and hinder downstream analytical tasks. Conventional baseline correction methods typically rely on iterative optimization with manual parameter tuning, rendering end-to-end automated processing difficult. Although deep learning frameworks have emerged to automate background removal, existing models often struggle to balance the preservation of sharp, localized Raman peaks with the accurate modeling of broad, slowly varying baselines. To overcome this limitation, we introduce DU-GridNet, a compact one-dimensional deep network utilizing a multi-resolution grid topology to automate Raman baseline correction. Through multi-scale feature propagation via interconnected sampling blocks, the model simultaneously isolates sharp Raman peaks and fits broad fluorescence backgrounds without manual intervention. Benchmark evaluations demonstrate that the proposed model achieves superior baseline estimation accuracy compared to existing methods. Furthermore, validation on empirical Raman datasets confirms its robust capacity to eliminate diverse fluorescence backgrounds while preserving critical spectral fidelity.
Related Concept Videos
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
NMR Spectrometers: Resolution and Error Correction
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
IR Spectroscopy: Hooke's Law Approximation of Molecular Vibration
According to Hooke's law, the vibrational frequency is directly proportional to the...
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are slanted or...