Pulse Wave Velocity Testing in the Baltimore Longitudinal Study of Aging
Published on: February 7, 2014
Assessment of the Variability of an Echo-Doppler Method for Measurement of Aortic Pulse Wave Velocity
Raffaele J Massarotto1,2, Timothy J Bradley3, James E Potts4
1Cardiovascular Physiology Laboratory, Faculty of Biomedical Physiology and Kinesiology, Simon Fraser University, Burnaby, British Columbia, Canada.
Background:
Pulse wave velocity (PWV) is an indirect measure of elastic arterial stiffness. Although various methods have been established to measure PWV, the Echo-Doppler method for measuring thoracic aortic PWV (a-PWV) is a highly available and cost-effective technique within the cardiology clinic. The purpose of this study was to assess the variability of a-PWV using the Echo-Doppler method and identify the optimal number of measurements providing the least variability of a-PWV in paediatric patients and healthy controls.
Methods:
Eighteen patients with a dilated proximal aorta and 21 healthy controls with a normal-sized proximal aorta (10-18 years) were evaluated. Imaging was performed in 4 studies across 2 visits. Variability comparisons were performed for short-term, medium-term, intra-reviewer, inter-reviewer, and inter-sonographer variability and analysed via intraclass correlation coefficient (ICC) and Bland-Altman plots.
Results:
The ICCs were highest for 10 beats. The combined group ICCs for the variability comparisons for 10 beats were short-term (0.743), medium-term (0.723) intra-reviewer (0.978), inter-reviewer (0.729), and inter-sonographer (0.823). Bland-Altman plots showed good agreement for all variability comparisons.
Conclusions:
This study has established that the measurement over 10 beats produces the lowest variability and is recommended when assessing a-PWV in both patients with a dilated proximal aorta and controls. Further, reproducibility remains high, even when accounting for sources of measurement variation, such as differences between sonographers, reviewers, and study visits.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation