Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
Published on: October 17, 2013
Effects of segment chemistry on mechanical performance and hemocompatibility in polyurethane heart valve scaffolds
Farhad Sadeghi1, Yasaman Zamani1, Mia Tacklind1
1Department of Biomedical Engineering, University of California, Irvine, USA. arashkh@uci.edu.
Abstract:
Polycarbonate-based and polysiloxane-modified thermoplastic polyurethanes (TPUs) are promising candidates for cardiovascular implants because of their tunable mechanical properties and favorable biocompatibility. In this study, we established comparative formulation-property-hemocompatibility criteria for polymeric heart valve scaffolds through a systematic evaluation of five commercial TPU formulations: carbothane AC-4095A (CB95-AC), carbothane AC-4075A (CB75-AC), carbothane PC-3575A (CB75-PC), chronoflex C65D (CF), and quadrasil ARCS 90A (Qsil). This comparative analysis provides practical material-selection criteria based on the balance between elastic recovery, tear resistance, oxidative stability, and hemocompatibility. CF and CB95-AC exhibited the highest stiffness and tensile strength, whereas CB75-AC and CB75-PC demonstrated greater elasticity and lower residual strain during cyclic loading. CB75-AC exhibited the lowest hysteresis and best elastic recovery; however, it also exhibited low tear resistance and slightly elevated hemolysis (∼4%), highlighting the tradeoff between elastic recovery and mechanical robustness. All TPU materials maintained platelet counts and partial thromboplastin time values within ASTM-defined acceptable ranges. FTIR analysis revealed formulation-dependent differences in urethane carbonyl hydrogen bonding and hard-domain association that were associated with differences in mechanical behavior and surface characteristics. DSC revealed formulation-dependent differences in thermal transitions, whereas WAXD showed broadly similar, predominantly amorphous structures with diffuse scattering centered at approximately 20-21° 2θ and no distinct crystalline reflections. Preliminary short-term implantation of a CB95-AC polymeric mitral valve in an ovine model showed no gross thrombotic attachment on the leaflet surfaces after 22 days. Overall, the results demonstrate formulation-dependent tradeoffs among elastic recovery, tear resistance, oxidative stability, and hemocompatibility in TPU-based heart valve scaffolds. These findings provide practical material-selection criteria for future polymeric heart valve scaffold development.
More Related Videos
10:15The Use of the Ex Vivo Chandler Loop Apparatus to Assess the Biocompatibility of Modified Polymeric Blood Conduits
Published on: August 20, 2014
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020