Evaluating Optimization-Based Image Refinement for 3-D Ultrasound Mosaicking: A Sensitivity Analysis of Clinical and Benchmark Data
Related Experiment Videos
Maria Antico1, Sarah A Mason2, Emma Harris2
1Australian E-health Research Centre, CSIRO Health and Biosecurity, Brisbane, QLD, Australia; Faculty of Health, Medical and Radiation Sciences, Queensland University of Technology, Brisbane, Queensland, Australia.
Objective:
This study evaluates the performance and sensitivity of an optimization-based registration framework for mosaicking partially overlapping 3-D ultrasound volumes to expand the clinical field of view. We specifically characterize the algorithm's ability to recover from varying degrees of initial tracking uncertainty across different anatomical contexts.
Methods:
Performance was assessed on (i) a synthetic benchmark knee data set (240 volume pairs from 16 volunteers) where controlled perturbations (±3 mm/°) were applied to a known ground-truth pose to evaluate convergence limits, and (ii) a clinical abdominal data set (128 volume pairs from 6 volunteers, 3-5 partially overlapping volumes per session), where the algorithm was initialized both at the reference optical tracking pose and under broader simulated tracking errors (up to ±10 mm/°).
Results:
On the synthetic data set, the method achieved high registration success (95.4%-96.2%) and accuracy (93.3%-93.8%). On the clinical data set, deviations from the reference increased with wider initial perturbations, highlighting a limited capture range. However, when initialized at the tracking-based position, the image-based refinement improved alignment compared to tracking alone in 44% of cases, with Dice gains up to 10%.
Conclusion:
These findings demonstrate that while image-based refinement can enhance hardware-based mosaicking, its reliability depends heavily on the accuracy of the initial coarse alignment and the search constraints applied.