Double-bowtie filter design for pediatric spectral CT imaging: A fan beam simulation study
Related Experiment Video
Updated: Nov 6, 2026

Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
Yinglin Ge1,2, Olivia F Sandvold1,2, Roland Proksa1
1Department of Radiology, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
Background:
Despite concerns about radiation dose in pediatric patients, computed tomography (CT) remains a first-line imaging modality for children. Spectral CT provides additional diagnostic information that benefits pediatric imaging. However, most existing spectral CT systems are primarily optimized for adult patients.
Purpose:
To develop and evaluate a novel double bowtie filter integrating a K-edge material layer with a conventional Teflon filter for CT. The proposed design aims to enhance spectral signal-to-noise ratio (SNR) and spectral separation within the projection domain while maintaining radiation dose levels suitable for pediatric imaging.
Methods:
A simulation framework was developed and used to model a rapid kVp-switching CT system operating at 70/110 kVp with realistic tube power and geometric constraints. Pediatric phantoms of three sizes (100-200 mm anterior-posterior width) were used to evaluate performance. Five accessible and safe filter materials-gadolinium (Gd), holmium (Ho), erbium (Er), silver (Ag), and tin (Sn)-were tested in combination with a Teflon bowtie. System performance was quantified using virtual monoenergetic image (VMI) SNR at 40 and 70 keV, and the area under the monoenergetic SNR curve (AUMC) as a comprehensive spectral image quality metric. Dose consistency was ensured using a traditional Teflon bowtie as the reference.
Results:
The Teflon + Gd configuration achieved the highest performance, improving mean AUMC over all fan angles by 53.6% on average and up to 56% for the largest phantom. VMI SNR increased by approximately 46% at 40 keV and 42% at 70 keV.
Conclusions:
The double-bowtie concept substantially enhances spectral performance. The Teflon + Gd design provides a manufacturable, pediatric-optimized solution adaptable to kVp-switching and other spectral CT architectures, offering improved diagnostic quality at low dose levels.