Case report: complete pilot balloon detachment in a ventilator-dependent patient with high-risk tracheostomy exchange
Related Experiment Video
Updated: Nov 9, 2026

A Novel Rescue Technique for Difficult Intubation and Difficult Ventilation
Published on: January 17, 2011
Xinyu Lai1, Yan Huang1,2, Mingyu Hu1
1Intensive Care Unit, Xiangya Changde Hospital, Changde, Hunan, China.
Background:
Complete detachment of a pilot balloon and one-way valve eliminates the standard route for tracheostomy-cuff inflation and pressure measurement. Definitive tube exchange may be hazardous in selected patients.
Case Description:
A 74-year-old ventilator-dependent woman with a long-term tracheostomy developed an audible upper-airway leak, a ventilator leak alarm, and reduced displayed tidal volume. Inspection showed complete loss of the pilot balloon and one-way valve, while the residual inflation line remained attached to the tube. Computed tomography documented post-tracheostomy changes but did not report peri-tracheostomy granulation. Otolaryngology bedside examination identified granulation tissue below the stoma and considered tube removal and exchange difficult, with risks of bleeding, aspiration, respiratory difficulty, and loss of the airway. After multidisciplinary assessment and informed shared decision-making, the existing tube was retained and an exceptional bridge manoeuvre was used to access the residual inflation-line lumen, inject approximately 3 mL of air, and occlude the line. Because the pilot balloon and one-way valve were absent, the standard connection for a cuff manometer no longer existed; intracuff pressure was not measured. The reported volume was case-specific, was not a target, and did not establish a safe cuff pressure, efficacy, or safety. Routine hourly observations near the event were not immediate pre-/post-intervention measurements and cannot demonstrate an intervention-related physiological effect. Intermittent leakage persisted, and the defective tube remained in situ at family-requested discharge.
Conclusion:
This case highlights recognition of cuff-system failure, specialist escalation, shared risk assessment, and surveillance when exchange is hazardous. This case-specific manoeuvre should not be interpreted as a recommended bedside procedure. Definitive management remains replacement of the defective tracheostomy tube whenever safely feasible.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Tracheostomy Suctioning II: Procedure
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask and...