Calibration of X-ray multimeters for radiography
Related Experiment Video
Updated: Nov 9, 2026

X-ray Beam Induced Current Measurements for Multi-Modal X-ray Microscopy of Solar Cells
Published on: August 20, 2019
Paula Toroi1, Viivi Valkama2, Touko Kaasalainen3
1STUK - Radiation and Nuclear Safety Authority, Vantaa, Finland; HUS Diagnostic Center, Radiology, Helsinki University and Helsinki University Hospital, Helsinki, Finland.
Introduction:
Accurate determination of air kerma, tube‑voltage‑related quantities (TV), and half‑value layer (HVL) is essential for reliable dosimetry in X‑ray imaging. X‑ray multimeters (XMMs) are widely used for these measurements; however, their performance depends on spectral characteristics and measurement conditions. This study evaluates the consistency and clinical applicability of calibration coefficients derived using industrial and clinical X‑ray systems across a range of clinically relevant radiation qualities.
Methods:
Eight XMMs from four manufacturers were calibrated for air kerma, TV, and HVL using one laboratory and three clinical systems. Reference values were established using traceable ionization chambers, high voltage dividers, and spectrometry. Calibration coefficients were assessed for standard aluminium filtered radiation qualities and for an extended range including 0-0.9 mm of copper filtration. Energy dependence, X‑ray system dependence, and the impact of applying calibration coefficients were evaluated.
Results:
For air kerma, calibration coefficients derived using industrial and clinical systems were highly consistent. Energy dependence variations were within 5% and could be reduced to below 1% by applying radiation‑quality‑specific calibration coefficients. TV measurements showed the strongest dependence on the X‑ray system; energy dependence variations were generally below 4% and could be improved only using clinical‑system‑based calibrations. For HVL measurements, energy dependence variations were typically within 6% and were reduced below 1% by calibration, except for two XMMs exhibiting system‑dependent behaviour.
Conclusions:
XMMs can provide reliable measurements across clinically relevant radiation qualities, including copper‑filtered beams, when radiation‑quality‑specific calibration coefficients are applied. Interpolation based on a single radiation quality parameter is insufficient; at least two parameters are required.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
X-ray Imaging
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Radiological Investigation I: X-ray and CT
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...