Stim-CODE: PNS and CNS Constraint Optimized Diffusion-Encoding for Neuroimaging on 200 mT/m Whole-Body Gradients
Related Experiment Video
Updated: Nov 11, 2026

Neuroimaging-Guided TMS–EEG for Real-Time Cortical Network Mapping
Published on: June 13, 2025
Ariel J Hannum1,2,3, Michael Loecher1,2, Qingping Chen4
1Department of Radiology, Stanford University, Stanford, California, USA.
Purpose:
Ultra-high-performance whole-body MRI gradients ( 200mT/m, 200 T/m/s) can reduce echo times (TE) in diffusion-weighted imaging (DWI), improving SNR. However, full hardware utilization is limited by cardiac (CNS) and peripheral (PNS) nerve stimulation thresholds. This work demonstrates Stimulation Constraint Optimized diffusion-encoding (Stim-CODE) for neuro diffusion.
Methods:
Stim-CODE uses the gradient optimization (GrOpt) toolbox to optimally design diffusion-encoding gradient waveforms subject to a CNS or PNS "constraint envelope" based on other gradient events in the sequence. Stim-CODE was evaluated in simulations and in vivo protocols with in-plane resolutions of 1.5 1.5 mm and 1.0 1.0 mm across G = [45, 80, 200] mT/m and T/m/s. Healthy volunteers (N = 10) underwent neuro DWI at b = 1000 s/mm with Monopolar, Stim-CODE- , Bipolar, and Stim-CODE- encoding. SNR, apparent diffusion coefficient (ADC) maps, and ADC uncertainty were quantified. One volunteer underwent DTI with the same encoding schemes acquired at = [1000, 3000]s/mm .
Results:
Stim-CODE reduced TEs across all -values, with larger gains for higher resolutions, greater G , and -nulled waveforms. Phantom and in vivo experiments confirmed equivalent diffusion-weighting with TE savings and SNR improvements. Stim-CODE- had significant SNR gains over Bipolar scans, but some differences in global ADC were observed. DTI metrics were similar across scans, with Stim-CODE showing lower uncertainty.
Conclusion:
Stim-CODE maximizes hardware capabilities to achieve time-optimized diffusion-encoding while respecting CNS and PNS, enabling similar diffusion-weighting as conventional approaches with reduced TE and higher SNR.