Evaluation of the Beckman creatinine analyzer.

J M Hicks, M Iosefsohn, S A Lewis

    Clinical Chemistry
    |June 1, 1979
    PubMed
    Summary

    The Beckman Creatinine Analyzer shows reliable results for serum and urine creatinine determination, comparable to other methods. Minor interference from acetone and acetoacetate was noted.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Mass transport of lipopolysaccharide induced H<sub>2</sub>O<sub>2</sub> detected by an intracellular carbon nanoelectrode sensor.

    Bioelectrochemistry (Amsterdam, Netherlands)·2020
    Same author

    Real-time bacterial detection with an intracellular ROS sensing platform.

    Biosensors & bioelectronics·2019
    Same author

    First Reports of Aphelenchoides fragariae on Royal Fern and on Hosta and Other Hosts in South Carolina.

    Plant disease·2019
    Same author

    First Report of Root-Knot Nematode Meloidogyne javanica on Chrysanthemum in Colombia.

    Plant disease·2019
    Same author

    Electrochemical communication with the inside of cells using micro-patterned vertical carbon nanofibre electrodes.

    Scientific reports·2016
    Same author

    A Case of Bilateral Cystic Partially Differentiated Nephroblastoma vs Cystic Wilms' Tumor: Highlighting a Diagnostic Dilemma.

    Urology·2016

    Area of Science:

    • Clinical Chemistry
    • Analytical Chemistry

    Background:

    • Accurate creatinine measurement is crucial for assessing kidney function.
    • Multiple analytical methods exist for serum and urine creatinine determination.

    Purpose of the Study:

    • To evaluate the performance of the Beckman Creatinine Analyzer.
    • To compare its results against the GEMSAEC Centrifugal Analyzer and a manual method.

    Main Methods:

    • Comparative analysis of creatinine determination in serum and urine.
    • Evaluation of linearity, stability, recovery, and precision.
    • Assessment of interference from common biological substances.

    Main Results:

    • The Beckman Creatinine Analyzer demonstrated comparable results to the GEMSAEC and manual methods.
    • Negligible interference was observed from bilirubin, protein, lipemia, uric acid, and hemolysis.
    • Significant interference was identified with acetone and acetoacetate.

    Conclusions:

    • The Beckman Creatinine Analyzer is a reliable tool for creatinine measurement.
    • Potential interference from ketones (acetone, acetoacetate) should be considered during analysis.