Visual and automated differential leukocyte counts. A comparison study of three instruments
American Journal of Clinical Pathology
|June 1, 1979
Summary
Automated leukocyte counting methods, including Larc, Hematrak, and Hermalog D, showed good accuracy. Cell distribution on blood smears significantly impacted pattern-recognition classifier variability.
Area of Science:
- Hematology
- Clinical Pathology
- Medical Technology
Background:
- Manual microscopic leukocyte counting is a standard diagnostic procedure.
- Automated methods offer potential for increased efficiency and standardization.
Purpose of the Study:
- To compare the accuracy and precision of two pattern-recognition leukocyte classifiers (Larc, Hematrak) and a cytochemical automated method (Hermalog D) against manual referee and technologist counts.
- To evaluate the impact of patient population and slide preparation on automated cell counting.
Main Methods:
- Comparison of cell counts from referee microscopy, multiple technologists, Larc, Hematrak, and Hermalog D.
- Testing with samples from random patients, selected abnormal patients, and healthy individuals.
- Analysis of factors contributing to count variability, including cell distribution on smears.
Main Results:
- All automated methods demonstrated good accuracy and effectively flagged abnormal results for review.
- Variability in pattern-recognition counts was primarily attributed to cell distribution on blood smears.
- Larc showed high sensitivity to cellular alterations, leading to slide rejections; Hemalog D exhibited superior precision.
Conclusions:
- Automated leukocyte counters are accurate and reliable for clinical use.
- Slide preparation techniques are critical for pattern-recognition instruments and influence instrument selection.
- Hermalog D offers the highest precision among the evaluated automated methods.