Related Experiment Videos
Class II (laminar flow) biological safety cabinet.
Journal of Clinical Pathology
|May 1, 1979
Summary
Microbiological safety cabinets in the UK require careful selection and monitoring. Most Class II cabinets tested showed poor aerosol containment, posing risks with pathogenic materials.
Area of Science:
- Microbiology
- Laboratory Safety Engineering
Background:
- Class II (laminar flow) safety cabinets are critical for laboratory containment.
- Previous assessments of their effectiveness in the UK are limited.
- Ensuring adequate biological containment is paramount for preventing aerosolized pathogen exposure.
Purpose of the Study:
- To survey the microbiological effectiveness of Class II safety cabinets used in the UK.
- To evaluate their aerosol containment capabilities compared to Class I cabinets.
- To inform purchasing and maintenance practices for safety cabinets.
Main Methods:
- A microbiological survey was conducted on nine Class II safety cabinets.
- Aerosol containment was assessed for each unit.
- Performance was benchmarked against Class I (exhaust protective) cabinets.
Main Results:
- Only two of the nine tested Class II cabinets demonstrated adequate aerosol containment.
- The majority of cabinets posed potential hazards when used with pathogenic materials.
- Performance varied significantly among the surveyed units.
Conclusions:
- Many Class II safety cabinets currently in use in the UK do not provide sufficient protection.
- New specifications from the National Sanitation Foundation and British Standards Institute are expected to improve cabinet performance.
- Careful cabinet selection, regular monitoring, and post-servicing checks are essential for maintaining safety.