Effect on pacemakers of airport weapons detectors
Abstract:
An investigation was carried out using a variety of pacemakers and all the types of weapons detectors in common use in Canada, to determine whether or not such detectors present a hazard to pacemaker bearers. The results indicate that only left-side implants of unipolar sensing pacemakers are likely to be affected, that ventricular fibrillation initiated by interference-induced competitive pacing is the only conceivable hazard, but that the probability of 10(-9) for the occurrence of this event is so low that it may be completely disregarded. Physicians may therefore reassure pacemaker bearers of their safety in and around airport weapons detectors.
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...