Effect of plate size and location of disc on zone diameter in the disc antimicrobial susceptibility test

Applied Microbiology
|September 1, 1971
PubMed

Insights

Smaller petri dishes (100-mm) and various dispensers are often used in antimicrobial susceptibility testing. However, these may lead to difficulties in zone diameter measurement, potentially impacting clinical laboratory quality.

Area of Science:

  • Clinical Microbiology
  • Antimicrobial Resistance
  • Laboratory Diagnostics

Background:

  • The standard antimicrobial disc susceptibility test utilizes 150-mm petri dishes.
  • Many laboratories opt for 100-mm plates and diverse mechanical dispensers for practical reasons.
  • The impact of these substitutions on test accuracy and precision remains a concern.

Purpose of the Study:

  • To compare zone diameters obtained using standard 150-mm plates versus smaller 100-mm plates with various dispensers.
  • To evaluate the precision and analytical variability of antimicrobial disc susceptibility testing with different plate sizes and dispensers.
  • To assess the potential impact on clinical laboratory quality when using non-standard equipment.

Main Methods:

  • Antimicrobial disc susceptibility testing was performed on Staphylococcus aureus and Escherichia coli.
  • Tests were conducted using both 150-mm and 100-mm petri dishes with different mechanical dispensers.
  • Zone diameters were measured and compared between the various experimental conditions.

Main Results:

  • Zone diameters were consistently smaller at the edges compared to the center of plates (0.4 to 1.2 mm difference).
  • Mean zone diameters showed minor differences (up to 1.2 mm) or no significant differences across plate sizes and dispensers.
  • All measurements fell within accepted analytical variability for the procedure; however, measuring zones at the periphery of smaller plates was more difficult.

Conclusions:

  • No specific combination of plates, dispensers, or disc locations demonstrated superior precision in zone diameter measurement.
  • The use of smaller plates may introduce practical challenges in zone measurement, potentially affecting the quality of work in busy clinical laboratories.
  • Further evaluation is needed to ensure the reliability and comparability of results when deviating from standard antimicrobial susceptibility testing protocols.