Some problems in the use of the Coulter counter for erythrocyte total counts and volume distribution

Normal erythrocytes were subject to swelling and some haemolysis when suspended in a NaCl-formaldehyde-Na(2)EDTA diluent of pH 4.9 for routine blood-cell counting in the Coulter counter. A phosphate-buffered saline of pH 7.4 was substituted and has proven satisfactory for total erythrocyte counts. Polyvinyltoluene latex was unsuitable for volume calibration of the counter, because of aggregation of the particles in saline. Ragweed pollen grains gave a calibration factor that agreed well with that for fresh erythrocytes. Recent batches of Dade's Erythro-Trol (a suspension of fixed erythrocytes) were satisfactory for both total counts and volume calibration.

Related Concept Videos

Flow Cytometry01:23

Flow Cytometry

The development of flow cytometry techniques began in 1934 with initial attempts by Andrew Moldavan, a bacteriologist who counted the cells in a flowing capillary system. Moldavan pumped cells through a capillary tube focused under a microscope for visualization. The invention of photometry allowed the measurement of differentially-stained cells, and Louis Kamentsky developed the first multiparameter flow cytometer in 1965 to identify and count the cancer cells in cervical tissue specimens.
In...
Coulometry: Overview01:00

Coulometry: Overview

Coulometry is one of the rapid, most accurate, and precise analytical techniques that determine the quantity of an analyte by measuring the electrical charge needed for its complete electrolysis without using any analytical standards. The total charge passed during electrolysis correlates with the analyte amount by Faraday's laws of electrolysis. For accurate coulometric measurements, a charge equal to Faraday's constant multiplied by the number of electrons involved in the relevant...
Controlled-Current Coulometry: Coulometric Titration01:18

Controlled-Current Coulometry: Coulometric Titration

Coulometric titrations are a form of titrimetric analysis where the reagent is generated electrically, and its amount is evaluated based on current and generating time. The electron serves as the standard reagent. The procedure is similar to conventional titrations, such as endpoint detection.
The fundamental requirements for coulometric titrations are (1) 100% efficiency in the reagent-generating electrode reaction and (2) a stoichiometric and preferably rapid reaction between the generated...
Controlled-Current Coulometry: Overview01:27

Controlled-Current Coulometry: Overview

Controlled current coulometry, also known as amperostatic coulometry, is a technique used in electrochemical analysis to measure the quantity of a substance through the controlled passage of current. It involves the application of a constant current to an electrochemical cell containing the analyte of interest. As the current flows through the cell, the analyte undergoes a redox reaction at the electrode surface, resulting in a charge transfer. By monitoring the time required for a certain...