Urinary inorganic phosphorus determinations
Related Experiment Videos
Journal of Clinical Pathology
|March 1, 1972
Abstract:
The Parekh-Jung method for determination of inorganic phosphorus in serum was applied to its determination in urine. Accuracy is good. The mean percentage recovery of added phosphorus was 100.2%. Forty analyses of a sample gave a mean value of 46.71 mg/100 ml +/- 0.76 standard deviation (between-batch precision). The relative usefulness of the Parekh-Jung method compared with those other methods is discussed.