Interference by Gastrografin with a spectrophotometric trypsin assay
Gut
|May 1, 1972
Summary
Gastrografin can interfere with pancreatic function tests. This contrast agent can cause falsely low trypsin values due to high absorbance and enzyme inhibition, impacting diagnostic accuracy.
Area of Science:
- Gastroenterology
- Biochemistry
- Diagnostic Assays
Background:
- Gastrografin is a contrast agent used in medical imaging.
- Accurate pancreatic function tests are crucial for diagnosing gastrointestinal disorders.
- Spectrophotometric assays are commonly used for enzyme determination.
Purpose of the Study:
- To investigate the potential interference of Gastrografin with spectrophotometric trypsin assays.
- To identify the mechanisms by which Gastrografin might affect assay results.
- To provide recommendations for optimizing pancreatic function testing.
Main Methods:
- Gastrografin was introduced into the intestinal tract.
- Trypsin values were determined using a spectrophotometric assay system.
- Interference was assessed by measuring absorbance at 254 nm and evaluating enzyme inhibition.
- Amylase and lipase determinations were also performed.
Main Results:
- Residual Gastrografin caused falsely low trypsin values in spectrophotometric assays.
- High absorbance of Gastrografin at 254 nm led to elevated background optical density.
- Gastrografin inhibited the esterase activity of the synthetic substrate used in the assay.
- No interference was observed with gelatin proteolysis by trypsin, amylase, or lipase determination.
Conclusions:
- Gastrografin interferes with spectrophotometric determination of trypsin.
- Avoid instilling Gastrografin before pancreatic function tests requiring trypsin measurement.
- The interference is attributed to absorbance and enzyme inhibition properties of Gastrografin.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
1.7K
Inductively coupled plasma–mass spectrometry (ICP–MS) is a highly selective and sensitive technique for accurate elemental analysis. Though the analysis of ICP–MS mass spectra is comparatively straightforward, it is affected by spectroscopic and non-spectroscopic interferences. Spectroscopic interferences arise when the plasma contains ionic species with an m/z value the same as the analyte ion. Spectroscopic interference can be categorized as isobaric, polyatomic ions, and...
1.7K
Atomic Absorption Spectroscopy: Interference
2.4K
Interference leads to systematic error in atomic absorption (AA) measurements by enhancing or diminishing the analytical signal or the background. These interferences can be grouped into three main categories: spectral interference, chemical interference, and physical interference.
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
2.4K
Automated Microbial Diagnostics
48
Automated diagnostic analyzers have transformed clinical microbiology by providing rapid and reliable methods for pathogen identification and antibiotic susceptibility testing. Among these systems, the Vitek 2 is widely used because it automates the traditionally labor-intensive processes of microbial identification (ID) and antibiotic susceptibility testing (AST), delivering standardized and timely results that are essential for effective patient care.Microbial Identification with ID CardsThe...
48