Whole-blood hemoglobin determinations: a comparison of methodologies
American Journal of Clinical Pathology
|December 1, 1979
Summary
The Coulter Model S accurately measures total hemoglobin (Hb) even with high carboxyhemoglobin (HbCO). Manual methods using less reliable diluents may yield higher results due to turbidity.
Area of Science:
- Clinical Chemistry
- Hematology
Background:
- Accurate hemoglobin (Hb) measurement is crucial for diagnosing anemia and monitoring treatment.
- Spectrophotometric methods for Hb determination can be affected by interfering substances like carboxyhemoglobin (HbCO) and sample turbidity.
- Variability in commercial standards and diluents impacts the reliability of manual Hb assays.
Purpose of the Study:
- To evaluate the accuracy of the Coulter Model S for total hemoglobin measurement in the presence of elevated carboxyhemoglobin (HbCO).
- To assess the quality and reliability of commercial diluents and standards used for spectrophotometric cyanmethemoglobin (HiCN) determinations.
- To identify optimal methods for manual hemoglobin measurement in clinical laboratories.
Main Methods:
- Comparison of hemoglobin measurements between the Coulter Model S and manual spectrophotometric methods.
- Testing of various commercial diluents and standards for cyanmethemoglobin (HiCN) assay performance.
- Evaluation of sample turbidity and carboxyhemoglobin (HbCO) interference on different assay methods.
- Assessment of the impact of centrifugation on diluent reliability.
Main Results:
- The Coulter Model S accurately quantifies total hemoglobin (Hb) content, unaffected by elevated carboxyhemoglobin (HbCO) levels.
- Commercial diluents exhibited variable quality, with many causing sample turbidity.
- Manual hemoglobin determinations were often higher than Coulter S measurements, potentially due to turbidity.
- The Van Kampen and Zijlstra reagent, enhanced by centrifugation, proved to be the most reliable diluent.
Conclusions:
- The Coulter Model S offers a reliable method for total hemoglobin determination, even with HbCO interference.
- Manual spectrophotometric Hb assays are susceptible to errors from diluent quality, turbidity, and HbCO.
- Narrow-band-pass spectrophotometers are not recommended for routine HiCN determinations due to sensitivity to turbidity and HbCO.