Evaluation of the Bactec serum gentamicin assay
Antimicrobial Agents and Chemotherapy
|November 1, 1979
Summary
A new radiometric assay for gentamicin is less precise than radioimmunoassay but acceptable when the latter is unavailable. This method offers an alternative for determining gentamicin levels in clinical laboratories.
Area of Science:
- Clinical Chemistry
- Analytical Chemistry
- Pharmacology
Background:
- Gentamicin is an aminoglycoside antibiotic requiring therapeutic drug monitoring.
- Radioimmunoassay (RIA) is an established method for quantifying gentamicin levels.
- Accurate gentamicin measurement is crucial for optimizing efficacy and minimizing toxicity.
Purpose of the Study:
- To compare the precision and correlation of a radiometric assay with a standard radioimmunoassay for gentamicin quantification.
- To evaluate the suitability of the radiometric assay as an alternative method.
Main Methods:
- A radiometric assay for gentamicin was developed and compared against a radioimmunoassay protocol.
- Precision was assessed using coefficients of variation (CV) for within-run, day-to-day, and overall runs.
- Correlation was determined by analyzing 84 patient serum samples using both methods.
Main Results:
- The radiometric method exhibited higher coefficients of variation for within-run, day-to-day, and overall runs compared to RIA.
- A high coefficient of correlation (0.82) was observed between the radiometric assay and RIA for patient sera.
- Despite lower precision, the radiometric method demonstrated acceptable performance.
Conclusions:
- The radiometric assay is a viable alternative for gentamicin level determination, particularly in settings lacking RIA.
- While less precise than RIA, its correlation suggests clinical utility.
- Laboratories can consider the radiometric method when RIA is not accessible for gentamicin monitoring.