Comparison of ascending and bracketing methods in pure tone audiometry. A multi-laboratory study
Scandinavian Audiology
|January 1, 1979
Summary
The ascending and bracketing methods for hearing threshold determination yield similar results and subject variability. However, the bracketing method takes significantly longer to complete than the ascending method.
Area of Science:
- Audiology
- Hearing Science
- Diagnostic Methods
Background:
- Accurate hearing threshold determination is crucial for diagnosing hearing loss.
- Manual audiometric methods, such as ascending and bracketing, are commonly used.
- Understanding the efficiency and comparability of these methods is important for clinical practice.
Purpose of the Study:
- To compare the ascending and bracketing methods for manual hearing threshold determination.
- To assess differences in mean thresholds, subject variability, and test time.
- To evaluate the consistency of results across multiple laboratories.
Main Methods:
- Four laboratories participated in the study.
- Hearing thresholds were manually determined using attenuators with a 5 dB step.
- Both ascending and bracketing methods were employed.
- A total of 214 ears were measured across various frequencies (0.5-6 kHz).
Main Results:
- No significant difference was found between mean hearing thresholds obtained with ascending and bracketing methods.
- The grand mean difference across all laboratories and frequencies was -0.31 dB.
- Subject variability was comparable between the two methods.
- The bracketing method required a significantly longer test duration compared to the ascending method.
Conclusions:
- Ascending and bracketing methods provide comparable hearing threshold data.
- The ascending method is more time-efficient for manual hearing threshold determination.
- Clinical audiology can utilize either method, with a consideration for testing time.