Related Experiment Videos
Evaluation of a modified high-performance liquid chromatography assay for acebutolol and its major metabolite
Abstract:
Extensive modification of an existing high-performance liquid chromatography assay for acebutolol and its major metabolite has markedly improved chromatographic stability eliminating the previous need for frequent adjustment of the eluent composition to accommodate continuous loss of column retention. The eluents now used and avoidance of the requirement for elevated column temperature may be significant factors in the ability to maintain column life over 8 months of continuous use with little decrease in retention. As a result of the improved chromatographic stability full advantage can now be taken of automatic injection devices for the unattended processing of large numbers of samples. A significant modification of the work-up of blood samples has improved precision of the assay in whole blood. Nevertheless, it is recommended that plasma samples rather than whole blood be analyzed, since the plasma assay is faster and still more precise.