Faulty anesthesia circuits: a source of environmental pollution in the operating room
Anesthesia and Analgesia
|May 1, 1977
Abstract:
Commonly used disposable anesthesia circuits were studied for leak and gas spillage. Trace anesthetic gas concentrations produced by these circuits in the anesthesiologist's breathing zone were analyzed by a Hewlett-Packard Gas Chromatograph. These measurements demonstrated ambient halothane (3.29 +/- 0.1 ppm) and N2O (333.5 +/- 2.31 ppm) concentrations well above target levels, when swivel-type disposable anesthesia circuits were used, despite the presence of standard gas-scavenging devices and appropriate operating room fresh air exchange rates. Lower ambient concentration levels (0.38 +/- 0.03 ppm halothane and 31.3 +/- 1.49 ppm N2O) were measured when Y-type disposable anesthesia circuits were used.