Standardized, automated procedure for measurement of serum hemolytic complement activity
Applied Microbiology
|November 1, 1968
Summary
This study introduces an automated method for measuring serum hemolytic complement activity (C') using an AutoAnalyzer. The new procedure standardizes complement fixation tests, allowing for rapid screening of up to 21 samples per hour.
Area of Science:
- Immunology
- Clinical Chemistry
Background:
- Complement fixation tests are crucial in diagnostics but often lack standardization and automation.
- Existing methods can be time-consuming and labor-intensive.
Purpose of the Study:
- To develop and validate an automated AutoAnalyzer manifold for complement fixation procedures.
- To standardize and streamline the measurement of serum hemolytic complement activity (C').
Main Methods:
- An AutoAnalyzer manifold was designed to mix reagents in proportions similar to the Laboratory Branch Complement Fixation test.
- The fixation stage was bypassed to directly measure serum hemolytic complement activity (C').
Main Results:
- The automated manifold effectively measured serum hemolytic complement activity (C').
- Up to 21 serum samples per hour could be screened using two appropriate dilutions.
- Normal adult C' titers measured by this automated procedure ranged from 58 to 103 C'H(50) units per ml.
Conclusions:
- The developed AutoAnalyzer manifold provides a standardized and automated approach for measuring serum hemolytic complement activity.
- This method offers efficient screening of complement activity, with potential applications in clinical diagnostics and research.