Automation of two-stage factor VIII assay
Thrombosis and Haemostasis
|April 30, 1978
Summary
Two automated methods for factor VIII (FVIII) assays, Depex and Electra, show comparable results for plasma samples. Electra is suitable for FVIII concentrates, unlike Depex which may yield inaccurate high values.
Area of Science:
- Clinical Chemistry
- Hematology
- Biotechnology
Background:
- Accurate quantification of factor VIII (FVIII) is crucial for diagnosing and managing bleeding disorders.
- Automated assays offer efficiency and reproducibility in clinical laboratories.
Purpose of the Study:
- To compare the performance of two automated two-stage factor VIII assays: Depex and Electra.
- To evaluate their suitability for routine clinical use, including assays of factor VIII concentrates.
Main Methods:
- The Depex method utilizes an electrical circuit completion for clot detection.
- The Electra method employs an optical density technique for clot detection.
- Both methods were compared using haemophilic and normal plasma samples, as well as factor VIII concentrates.
Main Results:
- Both Depex and Electra provided comparable factor VIII levels for haemophilic and normal plasmas.
- Both methods are considered valid for low and normal factor VIII levels.
- Electra is suitable for factor VIII concentrate assays, whereas Depex may produce falsely elevated results due to increased viscosity.
Conclusions:
- Both automated assays are reliable for plasma factor VIII determination.
- Electra demonstrates broader applicability, including accurate quantification of factor VIII concentrates.
- Depex's limitations with viscous samples necessitate careful consideration in concentrate analysis.