Related Experiment Videos
Suppression of a demand pacemaker in the presence of redundant transvenous right ventricular leads
Pacing and Clinical Electrophysiology : PACE
|January 1, 1980
Abstract:
A ventricular inhibited demand cardiac pacemaker was inhibited in the presence of intimate contact between the active temporary bipolar electrode ring and a retained inactive permanent bipolar electrode tip. Electromagnetic interference effects, lack of insulation, lead breaks, and loose connections were all ruled out as the cause of pacemaker suppression.